NBN EN 868-10:2009

Packaging for terminally sterilized medical devices - Part 10: Adhesive coated nonwoven materials of polyolefines - Requirements and test methods

WITHDRAWN

About this standard

Languages
German, English and French
Type
NBN
Standards committee
CEN/TC 102
Status
WITHDRAWN
Publication date
17 July 2009
ICS Code
11.080.30 (Sterilized packaging)
Withdrawn Date
30 January 2019

About this training

Summary

polyolefines, manufactured from nonwovens complying with EN 868-9 used for sterile barrier systems and/ or
packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of
use.
NOTE 1 The need for a protective packaging may be determined by the manufacturer and the user.
This part of EN 868 only introduces performance requirements and test methods that are specific to the
products covered by this part of EN 868 but does not add nor modify the general requirements specified in
EN ISO 11607-1.
As such, the particular requirements in 4.2 to 4.3 can be used to demonstrate compliance with one or more
but not all of the requirements of EN ISO 11607-1.
NOTE 2 When additional materials are used inside the sterile barrier system in order to ease the organization, drying
or aseptic presentation (e.g. inner wrap, container filter, indicators, packing lists, mats, instrument organizer sets, tray
liners or an additional envelope around the medical device) then other requirements, including the determination of the
acceptability of these materials during validation activities, may apply.
The materials specified in this part of EN 868 are intended for single use only.